INTENDED USE
HIV Rapid Test Cassette is a rapid chromatographic immunoassay for the qualitative detection of antibodies to Human Immunodeficiency Virus(HIV) in whole blood, serum or plasma to aid in the diagnosis of HIV infection.
INTRODUCTION
HIV stands for human immunodeficiency virus. HIV is a virus that harms your immune system by destroying a type of white blood cell that helps your body fight infection, making a person more vulnerable to other infections and diseases. It is spread by contact with certain bodily fluids of a person with HIV, most commonly during unprotected sex (sex without a condom or HIV medicine to prevent or treat HIV), or through sharing injection drug equipment.
AIDS(Acquired immunodeficiency syndrome) is the late stage of HIV infection that occurs when the body's immune system is badly damaged because of the virus.
Manufacturer | Dewei Medical Equipment Co., Ltd |
Product Categories | IVD Test |
Principle | Rapid Colloidal Gold Immunochromatography |
Detection | Qualitative |
Storage | 2~30 º C |
Component | Rapid Test |
Package | 25T/box, 50T/Box, 100T/Box |
Trademark | Dewei |
Origin | China |
MAIN CONTENTS
Test Cassette
Droppers
Buffer
Instructions for use
STORAGE AND STABILITY
Store at 2 ~ 30 º C in the sealed pouch for 24 months.
The test must remain in the sealed pouch until use.
Keep away from direct sunlight, moisture and heat.
Do not freeze.
TEST PROCEDURE
1. Remove the test from its sealed pouch, and place it on a clean, level surface. For best results, the assay should be performed within one hour.
2. Using the provided disposable pipette, transfer 3 drops of the specimen (approximately 75 µL) to each specimen well (S) of the device, then add one drop of buffer and start the timer.
3. Wait for the colored band(s) to appear. The result should be read at 15 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION
POSITIVE: Two colored bands appear on the membrane. One band appears in the control region (C) and another band appears in the test region (T).
NEGATIVE: Only one colored band appears, in the control region (C). No apparent colored band appears in the test region (T).
INVALID: Control band fails to appear. Results from any test which has not produced a control band at the specified read time must be discarded. Please review the procedure and repeat with a new test. If the problem persists, discontinue using the kit immediately and contact your local distributor.