INTENDED USE
The HBsAg Test (Serum/Plasma) is a rapid visual immunoassay for the qualitative, presumptive detection of HBsAg in human serum or plasma specimens. This kit is intended to be used as an aid in the diagnosis of HBV infection.
INTRODUCTION
Hepatitis B virus (HBV) is the prototypic member of the hepadnaviruses. Hepatitis B surface antigen (HBsAg) is located in the lipid envelope of this small DNA virus. During the replicative phase of the virus, this surface antigen is produced in excess and is detectable in the blood of the infected. The incubation period of HBV is 6 weeks to 6 months.
Manufacturer | Dewei Medical Equipment Co., Ltd |
Product Categories | IVD Test |
Principle | Rapid Colloidal Gold Immunochromatography |
Detection | Qualitative |
Storage | 2~30 º C |
Component | Rapid Test |
Package | 25T/box, 50T/Box, 100T/Box |
Trademark | Dewei |
Origin | China |
MAIN CONTENTS
Test Cassette
Droppers
Buffer
Instructions for use
STORAGE AND STABILITY
Store at 2 ~ 30 º C in the sealed pouch for 24 months.
Keep away from direct sunlight, moisture and heat.
Do not freeze.
TEST PROCEDURE
1. Remove the test from its sealed pouch, and place it on a clean, level surface. For best results, the assay should be performed within one hour.
2. Using the provided disposable pipette, transfer 3 drops of the specimen (approximately 75 µL) to each specimen well (S) of the device, then add one drop of buffer and start the timer.
3. Wait for the colored band(s) to appear. The result should be read at 15 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION
POSITIVE: Two colored bands appear on the membrane. One band appears in the control region (C) and another band appears in the test region (T).
NEGATIVE: Only one colored band appears, in the control region (C). No apparent colored band appears in the test region (T).
INVALID: Control band fails to appear. Results from any test which has not produced a control band at the specified read time must be discarded. Please review the procedure and repeat with a new test. If the problem persists, discontinue using the kit immediately and contact your local distributor.